Product Introduction
The ECO IC is an entry-level compact ion chromatograph from Metrohm (Model 2.925.0020), designed for routine analysis of anions, cations, organic acids and polar substances. The instrument comes standard with a conductivity detector, a Metrohm Suppressor Module (MSM) and MagIC Net chromatography workstation software, enabling a complete analysis workflow from separation to detection in a single system. It is an ideal choice for routine water quality analysis and university teaching.
Main Technical Specifications
- Model / Order No.: Eco IC (2.925.0020)
- Detector: Built-in conductivity detector
- Suppressor: Metrohm Suppressor Module (MSM)
- Chromatography workstation: MagIC Net Compact / Multi / Professional
- Sample loop: 10 µL for cation determination; 10 µL for suppressed anion determination; 100 µL for non-suppressed anion determination; 250 µL for MiPT partial loop injection; interchangeable depending on application
- Compatible columns: Metrosep A Supp 17 series (100/4.0, 150/4.0, 250/4.0 mm); other column models may also be used
- Recommended flow rate: 0.6 mL/min (Metrosep A Supp 17 series); maximum flow rate of 1.8 mL/min for A Supp 17-100/4.0
- Fluid path configuration: High-pressure pump, peristaltic pump (for suppressor regeneration), injection valve, inline filter, purge valve
- Intelligent functions: Automatic column chip recognition for reading column parameters and usage history; real-time monitoring of flow rate and backpressure, with statistics on injection count, operating time and service intervals
- Automation expansion: Can be equipped with the 863 Compact IC autosampler for fully automated analysis of up to 36 samples; supports Inline Ultrafiltration and eluent generation module
Product Features
- One system includes detector, suppressor and chromatography workstation, offering outstanding cost-effectiveness
- Suitable for routine analysis of anions and cations in water, organic acids and polar substances
- Self-monitoring intelligent design reduces the risk of operating errors and protects the column
- Can be upgraded to a fully automated analysis system at any time, reducing labor and time costs
- Made in Switzerland, with high hardware durability and stability
- Data management complies with GLP/FDA requirements
Technical Specifications
| Flow rate | 0.001 – 20 mL/min (0.01 – 5 mL/min with standard PEEK head) |
| Flow increment | 1 µL/min |
| Reproducibility of eluent flow | < 0.1% deviation |
| Pressure range (pump) | 0 – 50.0 MPa (0 – 500 bar) |
| Pressure range (pump head) | 0 – 35.0 MPa (0 – 350 bar) |
| Residual pulsation | < 1% |
| High-pressure pump | Serial dual-piston pump, self-optimizing |
| Injection valve max. pressure | 35 MPa (350 bar) |
| MSM solvent resistance | No restriction |
| Peristaltic pump speed | 0 – 42 rpm in 7 levels of 6 rpm |
| Peristaltic pumping | 0.3 mL/min at 18 rpm (standard tubing) |
| Automation | Up to 36 samples |
| Dimensions (W x H x D) | 262 x 468 x 362 mm |
| Weight | 14.8 kg (without accessories) |
| Ambient conditions | +5 to +45 °C, max. 80 % RH, non-condensing |
| Power supply | 100–240 V (±10 %), 50–60 Hz (±3 %) |
| Power consumption | 65 W typical, 25 W standby, 300 W max. |
Specifications compiled from the manufacturer's official documentation; the delivered configuration and its datasheet prevail.